Surface EMG classification for prosthesis control: fuzzy logic vs. artificial neural network

Electromyography control system (ECS) is a well-known technique for prosthesis control application. It consists of two main modules namely feature extraction and classification. This paper presents the investigation of the classification module in the ECS. The surface electromyographic (EMG) signal...

وصف كامل

محفوظ في:
التفاصيل البيبلوغرافية
المؤلفون الرئيسيون: Ahmad, Siti Anom, Khalid, Mohd Asyraf, Ishak, Asnor Juraiza, Md. Ali, Sawal Hamid
التنسيق: Conference or Workshop Item
اللغة:English
منشور في: SciTePress 2012
الوصول للمادة أونلاين:http://psasir.upm.edu.my/id/eprint/31662/1/31662.pdf
http://psasir.upm.edu.my/id/eprint/31662/
الوسوم: إضافة وسم
لا توجد وسوم, كن أول من يضع وسما على هذه التسجيلة!
الوصف
الملخص:Electromyography control system (ECS) is a well-known technique for prosthesis control application. It consists of two main modules namely feature extraction and classification. This paper presents the investigation of the classification module in the ECS. The surface electromyographic (EMG) signals were recorded from flexor and extensor muscles of the forearm during wrist flexion and extension. Standard deviation and mean absolute value were used to extract information from the raw EMG signals. Two different classifiers, fuzzy logic and artificial neural network were used in investigating the surface EMG signals. The classifier is responsible to determine the movement of the subject’s limb during specific moment. The two classifiers were compared in terms of their performance.