Characterization of enzymatic glucose biosensor in buffer solution, in artificial saliva, and in potassium ferricyanide by linear sweep voltammetry
Existing practice of glucose-level monitoring for diabetic patients involves frequent finger pricking for blood samples. This is not only constantly painful, but could make patients susceptible to infections. A non-invasive, portable, graphene-based biosensor is being developed to measure glucose l...
Saved in:
Main Authors: | , , |
---|---|
Format: | Conference or Workshop Item |
Language: | English English |
Published: |
AIP
2018
|
Subjects: | |
Online Access: | http://irep.iium.edu.my/68443/1/ISTEC%202018%20AIP%20Conference%20Proceeding%20Nurul%20Izzati%20et%20al%202018.pdf http://irep.iium.edu.my/68443/7/68443%20Characterization%20of%20Enzymatic%20Glucose%20Biosensor%20in%20Buffer.pdf http://irep.iium.edu.my/68443/ https://aip.scitation.org/doi/10.1063/1.5066993 |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
Summary: | Existing practice of glucose-level monitoring for diabetic patients involves frequent finger pricking for blood
samples. This is not only constantly painful, but could make patients susceptible to infections. A non-invasive, portable, graphene-based biosensor is being developed to measure glucose levels in saliva instead of in blood, by incorporating rGO-PEDOT:PSS composites as the transducer layer on a working electrode – a screen-printed carbon electrode (SPCE) 2 mm in diameter. To evaluate the performance of the biosensor, the electrochemical technique of linear sweep voltammetry (LSV) was used. The results showed sensitivity of rGO-PEDOT:PSS-GOx/SPCE in artificial saliva: 105 μA/mM (σ = 12.52) at no glucose to 0.2 mM glucose concentrations and 35.33 μA/mM (σ = 5.79) at glucose concentrations of 0.2 mM to 0.6 mM, with correlation coefficient of 0.99. The sensitivity of the biosensor in 10 mM PBS (pH 6.8) is 9.88 μA/mM (σ = 2.74), while in 100 mM potassium ferricyanide it is 10.48 μA/mM (σ = 0.49). The high sensitivity obtained in artificial saliva suggests that the rGO-PEDOT:PSS composite is suitable for use as a transducer layer for non-invasive glucose monitoring. |
---|