Preparation, scratch adhesion and anti-corrosion performance of TiO2-MgO-BHA coating on Ti6Al4V implant by plasma electrolytic oxidation technique

Bovine hydroxyapatite (BHA) (from cortical bone), was selected as the main electrolyte for plasma electrolytic oxidation (PEO) on Ti6Al4V implant. The prepared PEO coatings were examined by X-ray diffraction, field emission scanning electron microscope and energy-dispersive X-ray spectroscopy. The s...