Development and validation of a rapid ultra high performance liquid chromatography with tandem mass spectrometry method for the simultaneous determination of darunavir, ritonavir, and tenofovir in human plasma: Application to human pharmacokinetics

A sensitive ultra high performance liquid chromatography with tandem mass spectrometry method was developed for the simultaneous determination of darunavir, ritonavir and tenofovir in human plasma. Sample preparation involved a simple liquid-liquid extraction using 200 μL of human plasma extracted w...

Full description

Saved in:
Bibliographic Details
Main Authors: Reddy, A. V. B., Jaafar, Jafariah, Aris, A. B., Majid, Z. A., Umar, K., Talib, J., Madhavi, G.
Format: Article
Published: Wiley-VCH Verlag 2015
Subjects:
Online Access:http://eprints.utm.my/id/eprint/58279/
https://doi.org/10.1002/jssc.201500250
Tags: Add Tag
No Tags, Be the first to tag this record!
id my.utm.58279
record_format eprints
spelling my.utm.582792022-01-26T02:31:03Z http://eprints.utm.my/id/eprint/58279/ Development and validation of a rapid ultra high performance liquid chromatography with tandem mass spectrometry method for the simultaneous determination of darunavir, ritonavir, and tenofovir in human plasma: Application to human pharmacokinetics Reddy, A. V. B. Jaafar, Jafariah Aris, A. B. Majid, Z. A. Umar, K. Talib, J. Madhavi, G. Q Science (General) A sensitive ultra high performance liquid chromatography with tandem mass spectrometry method was developed for the simultaneous determination of darunavir, ritonavir and tenofovir in human plasma. Sample preparation involved a simple liquid-liquid extraction using 200 μL of human plasma extracted with methyl tert-butyl ether for three analytes and internal standard. The separation was accomplished on an Acquity UPLC BEH C18 (50 mm x 2.1 mm, 1.7 μm) analytical column using gradient elution of acetonitrile/methanol (80:20, v/v) and 5.0 mM ammonium acetate containing 0.01% formic acid at a flow rate of 0.4 mL/min. The linearity of the method ranged between 20.0 and 12 000 ng/mL for darunavir, 2.0 and 2280 ng/mL for ritonavir, and 14.0 and 1600 ng/mL for tenofovir using 200 μL of plasma. The method was completely validated for its selectivity, sensitivity, linearity, precision and accuracy, recovery, matrix effect, stability, and dilution integrity. The extraction recoveries were consistent and ranged between 79.91 and 90.04% for all three analytes and internal standard. The method exhibited good intra-day and inter-day precision between 1.78 and 6.27%. Finally the method was successfully applied for human pharmacokinetic study in eight healthy male volunteers after the oral administration of 600 mg darunavir along with 100 mg ritonavir and 100 mg tenofovir as boosters. Wiley-VCH Verlag 2015 Article PeerReviewed Reddy, A. V. B. and Jaafar, Jafariah and Aris, A. B. and Majid, Z. A. and Umar, K. and Talib, J. and Madhavi, G. (2015) Development and validation of a rapid ultra high performance liquid chromatography with tandem mass spectrometry method for the simultaneous determination of darunavir, ritonavir, and tenofovir in human plasma: Application to human pharmacokinetics. Journal Of Separation Science, 38 (15). pp. 2580-2587. ISSN 1615-9306 https://doi.org/10.1002/jssc.201500250 DOI: 10.1002/jssc.201500250
institution Universiti Teknologi Malaysia
building UTM Library
collection Institutional Repository
continent Asia
country Malaysia
content_provider Universiti Teknologi Malaysia
content_source UTM Institutional Repository
url_provider http://eprints.utm.my/
topic Q Science (General)
spellingShingle Q Science (General)
Reddy, A. V. B.
Jaafar, Jafariah
Aris, A. B.
Majid, Z. A.
Umar, K.
Talib, J.
Madhavi, G.
Development and validation of a rapid ultra high performance liquid chromatography with tandem mass spectrometry method for the simultaneous determination of darunavir, ritonavir, and tenofovir in human plasma: Application to human pharmacokinetics
description A sensitive ultra high performance liquid chromatography with tandem mass spectrometry method was developed for the simultaneous determination of darunavir, ritonavir and tenofovir in human plasma. Sample preparation involved a simple liquid-liquid extraction using 200 μL of human plasma extracted with methyl tert-butyl ether for three analytes and internal standard. The separation was accomplished on an Acquity UPLC BEH C18 (50 mm x 2.1 mm, 1.7 μm) analytical column using gradient elution of acetonitrile/methanol (80:20, v/v) and 5.0 mM ammonium acetate containing 0.01% formic acid at a flow rate of 0.4 mL/min. The linearity of the method ranged between 20.0 and 12 000 ng/mL for darunavir, 2.0 and 2280 ng/mL for ritonavir, and 14.0 and 1600 ng/mL for tenofovir using 200 μL of plasma. The method was completely validated for its selectivity, sensitivity, linearity, precision and accuracy, recovery, matrix effect, stability, and dilution integrity. The extraction recoveries were consistent and ranged between 79.91 and 90.04% for all three analytes and internal standard. The method exhibited good intra-day and inter-day precision between 1.78 and 6.27%. Finally the method was successfully applied for human pharmacokinetic study in eight healthy male volunteers after the oral administration of 600 mg darunavir along with 100 mg ritonavir and 100 mg tenofovir as boosters.
format Article
author Reddy, A. V. B.
Jaafar, Jafariah
Aris, A. B.
Majid, Z. A.
Umar, K.
Talib, J.
Madhavi, G.
author_facet Reddy, A. V. B.
Jaafar, Jafariah
Aris, A. B.
Majid, Z. A.
Umar, K.
Talib, J.
Madhavi, G.
author_sort Reddy, A. V. B.
title Development and validation of a rapid ultra high performance liquid chromatography with tandem mass spectrometry method for the simultaneous determination of darunavir, ritonavir, and tenofovir in human plasma: Application to human pharmacokinetics
title_short Development and validation of a rapid ultra high performance liquid chromatography with tandem mass spectrometry method for the simultaneous determination of darunavir, ritonavir, and tenofovir in human plasma: Application to human pharmacokinetics
title_full Development and validation of a rapid ultra high performance liquid chromatography with tandem mass spectrometry method for the simultaneous determination of darunavir, ritonavir, and tenofovir in human plasma: Application to human pharmacokinetics
title_fullStr Development and validation of a rapid ultra high performance liquid chromatography with tandem mass spectrometry method for the simultaneous determination of darunavir, ritonavir, and tenofovir in human plasma: Application to human pharmacokinetics
title_full_unstemmed Development and validation of a rapid ultra high performance liquid chromatography with tandem mass spectrometry method for the simultaneous determination of darunavir, ritonavir, and tenofovir in human plasma: Application to human pharmacokinetics
title_sort development and validation of a rapid ultra high performance liquid chromatography with tandem mass spectrometry method for the simultaneous determination of darunavir, ritonavir, and tenofovir in human plasma: application to human pharmacokinetics
publisher Wiley-VCH Verlag
publishDate 2015
url http://eprints.utm.my/id/eprint/58279/
https://doi.org/10.1002/jssc.201500250
_version_ 1724073238173057024