Design analysis of TMJ implant under physiological loading condition

The aim of this study was to determine the behavior of Temporomandibular joint implant made of different materials when it is subjected to loads resulted of daily movements. A three dimensional model of a lower mandible of a human were developed from a CT scan dataset. A commercial Temporomandibular...

全面介紹

Saved in:
書目詳細資料
Main Authors: Arabshahi, Zohreh, Kashani, Jamal, Rahimian Koloor, Seyed Saeid, Abdul Kadir, Mohammed Rafiq, Azari, Abbas
格式: Article
出版: 2012
主題:
在線閱讀:http://eprints.utm.my/id/eprint/46779/
http://dx.doi.org/10.4028/www.scientific.net/AMR.488-489.996
標簽: 添加標簽
沒有標簽, 成為第一個標記此記錄!
實物特徵
總結:The aim of this study was to determine the behavior of Temporomandibular joint implant made of different materials when it is subjected to loads resulted of daily movements. A three dimensional model of a lower mandible of a human were developed from a CT scan dataset. A commercial Temporomandibular joint implant and fixation screws were modeled. Three dimensional finite element models of implanted mandible analyzed under static loading of five clunching tasks. The investigation went to study on two materials of titanium and Cobalt-chromium alloy for implant. The equivalent strain of whole structure was computed and utilized for evaluation of implant design. The safety factor results showed that both materials were capable to carry the applied complex loads.