Characterization of EBT3 radiochromic film for precise measurement of output factors in stereotactic radiosurgery
Radiochromic film become a demanding tool in a work of characterizing dosimetric properties in small field stereotactic radiosurgery. The aim of this project is to perform a comparative study in measuring output factor using EBT3 radiochromic film and PTW PinPoint Ion Chamber. Output factor measu...
Saved in:
Main Author: | |
---|---|
Format: | Monograph |
Language: | English |
Published: |
Pusat Pengajian Sains Kesihatan, Universiti Sains Malaysia
2015
|
Subjects: | |
Online Access: | http://eprints.usm.my/58209/1/NUR%20ZARIFHA%20BINTI%20ZAINOL%20ALAM%20-%20e.pdf http://eprints.usm.my/58209/ |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
id |
my.usm.eprints.58209 |
---|---|
record_format |
eprints |
spelling |
my.usm.eprints.58209 http://eprints.usm.my/58209/ Characterization of EBT3 radiochromic film for precise measurement of output factors in stereotactic radiosurgery Alam, Nur Zarifha Zainol RD520-599.5 Surgery by region, system, or organ Radiochromic film become a demanding tool in a work of characterizing dosimetric properties in small field stereotactic radiosurgery. The aim of this project is to perform a comparative study in measuring output factor using EBT3 radiochromic film and PTW PinPoint Ion Chamber. Output factor measurements using EBT3 films and PTW PinPoint Ion Chamber were performed using 6MV photon beam from medical linear accelerator .The measurements was done in solid water phantom and CNMC Model 74-320 water phantom. Measurements were done at dmax (1.5cm ) depth inside water phantom and 100 cm source surface distance. Results: Percentage difference of measured output factor with EBT3 Gafchromic film and PTW Ion PinPoint Chamber was approximately 10-11%. Analysis on dose response for EBT3 and EBT2 gives difference approximately 4% in magnitude. The result obtained for scanning orientation study show minimal difference in output factors for different side orientation which differs by less than 2%. Variations for all doses were found to be less than 0.16 (net difference of optical density) for 4°C and 28°C but 0.026 for 37°C. Net optical density for the films stored in 37°C recorded huge difference in relative standard deviation which approximately only 4% compared to 4°C and 28°C which are 34-37% but still represents linear relationship with dose. EBT3 is suitable for measurements of output factor in stereotactic radiosurgery and can be use on validation of dose when high spatial resolution is required for clinical practice. Pusat Pengajian Sains Kesihatan, Universiti Sains Malaysia 2015 Monograph NonPeerReviewed application/pdf en http://eprints.usm.my/58209/1/NUR%20ZARIFHA%20BINTI%20ZAINOL%20ALAM%20-%20e.pdf Alam, Nur Zarifha Zainol (2015) Characterization of EBT3 radiochromic film for precise measurement of output factors in stereotactic radiosurgery. Project Report. Pusat Pengajian Sains Kesihatan, Universiti Sains Malaysia. (Submitted) |
institution |
Universiti Sains Malaysia |
building |
Hamzah Sendut Library |
collection |
Institutional Repository |
continent |
Asia |
country |
Malaysia |
content_provider |
Universiti Sains Malaysia |
content_source |
USM Institutional Repository |
url_provider |
http://eprints.usm.my/ |
language |
English |
topic |
RD520-599.5 Surgery by region, system, or organ |
spellingShingle |
RD520-599.5 Surgery by region, system, or organ Alam, Nur Zarifha Zainol Characterization of EBT3 radiochromic film for precise measurement of output factors in stereotactic radiosurgery |
description |
Radiochromic film become a demanding tool in a work of
characterizing dosimetric properties in small field stereotactic radiosurgery. The
aim of this project is to perform a comparative study in measuring output factor
using EBT3 radiochromic film and PTW PinPoint Ion Chamber. Output factor measurements using EBT3 films and PTW
PinPoint Ion Chamber were performed using 6MV photon beam from medical
linear accelerator .The measurements was done in solid water phantom and
CNMC Model 74-320 water phantom. Measurements were done at dmax (1.5cm )
depth inside water phantom and 100 cm source surface distance.
Results: Percentage difference of measured output factor with EBT3 Gafchromic film and PTW Ion PinPoint Chamber was approximately 10-11%.
Analysis on dose response for EBT3 and EBT2 gives difference approximately
4% in magnitude. The result obtained for scanning orientation study show
minimal difference in output factors for different side orientation which differs by
less than 2%. Variations for all doses were found to be less than 0.16 (net
difference of optical density) for 4°C and 28°C but 0.026 for 37°C. Net optical
density for the films stored in 37°C recorded huge difference in relative standard
deviation which approximately only 4% compared to 4°C and 28°C which are
34-37% but still represents linear relationship with dose. EBT3 is suitable for measurements of output factor in stereotactic
radiosurgery and can be use on validation of dose when high spatial resolution
is required for clinical practice. |
format |
Monograph |
author |
Alam, Nur Zarifha Zainol |
author_facet |
Alam, Nur Zarifha Zainol |
author_sort |
Alam, Nur Zarifha Zainol |
title |
Characterization of EBT3 radiochromic film for
precise measurement of output factors in
stereotactic radiosurgery |
title_short |
Characterization of EBT3 radiochromic film for
precise measurement of output factors in
stereotactic radiosurgery |
title_full |
Characterization of EBT3 radiochromic film for
precise measurement of output factors in
stereotactic radiosurgery |
title_fullStr |
Characterization of EBT3 radiochromic film for
precise measurement of output factors in
stereotactic radiosurgery |
title_full_unstemmed |
Characterization of EBT3 radiochromic film for
precise measurement of output factors in
stereotactic radiosurgery |
title_sort |
characterization of ebt3 radiochromic film for
precise measurement of output factors in
stereotactic radiosurgery |
publisher |
Pusat Pengajian Sains Kesihatan, Universiti Sains Malaysia |
publishDate |
2015 |
url |
http://eprints.usm.my/58209/1/NUR%20ZARIFHA%20BINTI%20ZAINOL%20ALAM%20-%20e.pdf http://eprints.usm.my/58209/ |
_version_ |
1765297662571053056 |
score |
13.211869 |