Multivariate optimisation and validation of the analytical GC-FID for evaluating organic solvents in radiopharmaceutical
Objective: Analytical gas chromatography in line with a flame ionisation detector (GC-FID) method was developed and validated for evaluating organic solvents in radiopharmaceutical samples [18F]fluoroethyl-tyrosine ([18F]FET [18F] fluoromisonidazole ([18F]FMISO) and [18F] fluorothymidine ([18F]FLT)....
Saved in:
Main Authors: | , , , , |
---|---|
Format: | Article |
Published: |
Elsevier
2021
|
Online Access: | http://psasir.upm.edu.my/id/eprint/94303/ https://www.sciencedirect.com/science/article/pii/S1018364721002159?via%3Dihub |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
Summary: | Objective: Analytical gas chromatography in line with a flame ionisation detector (GC-FID) method was developed and validated for evaluating organic solvents in radiopharmaceutical samples [18F]fluoroethyl-tyrosine ([18F]FET [18F] fluoromisonidazole ([18F]FMISO) and [18F] fluorothymidine ([18F]FLT). Variables of the carrier gas flow (mL min-1 ) and a split ratio of injection on the response of analysis time and resolution were optimised with the assistance of a two-level full factorial design and desirability function of Derringer.
Methods: The proposed procedure was validated following the International Council for Harmonisation of Technical Requirements for Pharmaceuticals for Human Use (ICH) Q2 (R1) guideline.
Results: Excellent linearity, R2 > 0.990 indicated that approximately 99% of the variance in the response could be predicted from ethanol and acetonitrile concentrations ranging from 0.8 to 7.5 mg mL-1 and 0.1 to 1.0 mg mL-1, respectively. The proposed procedure proved to be selective, sensitive and accurate (85 to 105%), with excellent repeatability and precision (relative standard deviation (RSD) < 2%). In assessing the
robustness of the method, the proposed procedure also proved to be robust as the standardised effects values (SE) were insignificant (p > 0.05).
Conclusion: The proposed method has also been successfully used for the quantitative determination of ethanol and acetonitrile in [18F]FET, [18F]FMISO and [18F]FLT samples. Analytes were well resolved (R, 7.9 – 8.1) within 3.5 min even though the column had a larger, 0.53 mm internal diameter. The proposed method is, therefore, relevant for routine organic solvent quality control analysis of any 18Fradiopharmaceutical derivatives. |
---|