Front end of low power transmitter for wireless capsule endoscope

A low power transmitter front end for the application of a capsule endoscope operating at ISM band of 434 MHz has been proposed. The proposed architecture is aim to achieve low power consumption which is the essential parameter in a capsule endoscope. The transmitter front end is implemented us...

全面介紹

Saved in:
書目詳細資料
Main Authors: Lioe, De Xing, Shafie, Suhaidi, Ramiah, Harikrishnan, Sulaiman, Nasri
格式: Conference or Workshop Item
出版: IEEE
在線閱讀:http://psasir.upm.edu.my/id/eprint/39267/
標簽: 添加標簽
沒有標簽, 成為第一個標記此記錄!
實物特徵
總結:A low power transmitter front end for the application of a capsule endoscope operating at ISM band of 434 MHz has been proposed. The proposed architecture is aim to achieve low power consumption which is the essential parameter in a capsule endoscope. The transmitter front end is implemented using Silterra 0.13 um CMOS technology and consists of an upconversion mixer and a voltage controlled oscillator. The current consumption of the system is 1.57 mA with the supply voltage of 1.2 V. Output power is found to be -14.4 dBm and the phase noise of the ring oscillator is -81.76 dBc/Hz at offset frequency of 160 kHz.