In-silico Analysis of In-stent Restenosis Prediction of Stented Renal Artery
In-stent restenosis (ISR) is a common complication after renal artery stenting, and stent design may influence the formation of ISR. This study aims to predict the effect of stent design on the risk of ISR formation after renal artery stenting. Using computer-aided design, a simplified renal artery...
Saved in:
Main Authors: | , , , , , |
---|---|
Format: | Article |
Language: | English |
Published: |
Universiti Teknologi MARA (UiTM), Malaysia
2024
|
Subjects: | |
Online Access: | http://ir.unimas.my/id/eprint/46724/1/2.JMECHE-2023-0210.pdf http://ir.unimas.my/id/eprint/46724/ https://jmeche.uitm.edu.my/browse-journals/special-issues/special-issue-2024-vol-13/ https://doi.org/10.24191/jmeche.v13i1.3754 |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
id |
my.unimas.ir-46724 |
---|---|
record_format |
eprints |
spelling |
my.unimas.ir-467242024-11-27T01:46:11Z http://ir.unimas.my/id/eprint/46724/ In-silico Analysis of In-stent Restenosis Prediction of Stented Renal Artery Nurin Tihani, Nazemi Muhammad Faiz, Md Shakhih Mohamad Ikhwan, Kori Rudiyanto, Philman Jong Kahar, Osman Ishkrizat, Taib RZ Other systems of medicine TJ Mechanical engineering and machinery In-stent restenosis (ISR) is a common complication after renal artery stenting, and stent design may influence the formation of ISR. This study aims to predict the effect of stent design on the risk of ISR formation after renal artery stenting. Using computer-aided design, a simplified renal artery model was modelled and later assembled with stents similar to commercial stents. Computational Fluid Mechanics (CFD) analysis was employed to simulate the blood flow in these stented models. Stent 1 exhibited superior hemodynamic performance with minimal low Wall Shear Stress (WSS) exposure in the stented region. Additionally, the study found that the thinnest Stent 1 was the most effective in reducing the risk of ISR formation. Thus, stent strut configuration with large spacing may reduce ISR risk, with the thinnest Stent 1 design showed promising performance in mitigating ISR. Hence, it is worth further investigation. Universiti Teknologi MARA (UiTM), Malaysia 2024-11-15 Article PeerReviewed text en http://ir.unimas.my/id/eprint/46724/1/2.JMECHE-2023-0210.pdf Nurin Tihani, Nazemi and Muhammad Faiz, Md Shakhih and Mohamad Ikhwan, Kori and Rudiyanto, Philman Jong and Kahar, Osman and Ishkrizat, Taib (2024) In-silico Analysis of In-stent Restenosis Prediction of Stented Renal Artery. Journal of Mechanical Engineering, SI (13). pp. 17-30. ISSN 2550-164X https://jmeche.uitm.edu.my/browse-journals/special-issues/special-issue-2024-vol-13/ https://doi.org/10.24191/jmeche.v13i1.3754 |
institution |
Universiti Malaysia Sarawak |
building |
Centre for Academic Information Services (CAIS) |
collection |
Institutional Repository |
continent |
Asia |
country |
Malaysia |
content_provider |
Universiti Malaysia Sarawak |
content_source |
UNIMAS Institutional Repository |
url_provider |
http://ir.unimas.my/ |
language |
English |
topic |
RZ Other systems of medicine TJ Mechanical engineering and machinery |
spellingShingle |
RZ Other systems of medicine TJ Mechanical engineering and machinery Nurin Tihani, Nazemi Muhammad Faiz, Md Shakhih Mohamad Ikhwan, Kori Rudiyanto, Philman Jong Kahar, Osman Ishkrizat, Taib In-silico Analysis of In-stent Restenosis Prediction of Stented Renal Artery |
description |
In-stent restenosis (ISR) is a common complication after renal artery stenting, and stent design may influence the formation of ISR. This study aims to predict the effect of stent design on the risk of ISR formation after renal artery stenting. Using computer-aided design, a simplified renal artery model was modelled and later assembled with stents similar to commercial stents. Computational Fluid Mechanics (CFD) analysis was employed to simulate the blood flow in these stented models. Stent 1 exhibited superior hemodynamic performance with minimal low Wall Shear Stress (WSS) exposure in the stented region. Additionally, the study found that the thinnest Stent 1 was the most effective in reducing the risk of ISR formation. Thus, stent strut configuration with large spacing may reduce ISR risk, with the thinnest Stent 1 design showed promising performance in mitigating ISR. Hence, it is worth further investigation. |
format |
Article |
author |
Nurin Tihani, Nazemi Muhammad Faiz, Md Shakhih Mohamad Ikhwan, Kori Rudiyanto, Philman Jong Kahar, Osman Ishkrizat, Taib |
author_facet |
Nurin Tihani, Nazemi Muhammad Faiz, Md Shakhih Mohamad Ikhwan, Kori Rudiyanto, Philman Jong Kahar, Osman Ishkrizat, Taib |
author_sort |
Nurin Tihani, Nazemi |
title |
In-silico Analysis of In-stent Restenosis Prediction of Stented Renal Artery |
title_short |
In-silico Analysis of In-stent Restenosis Prediction of Stented Renal Artery |
title_full |
In-silico Analysis of In-stent Restenosis Prediction of Stented Renal Artery |
title_fullStr |
In-silico Analysis of In-stent Restenosis Prediction of Stented Renal Artery |
title_full_unstemmed |
In-silico Analysis of In-stent Restenosis Prediction of Stented Renal Artery |
title_sort |
in-silico analysis of in-stent restenosis prediction of stented renal artery |
publisher |
Universiti Teknologi MARA (UiTM), Malaysia |
publishDate |
2024 |
url |
http://ir.unimas.my/id/eprint/46724/1/2.JMECHE-2023-0210.pdf http://ir.unimas.my/id/eprint/46724/ https://jmeche.uitm.edu.my/browse-journals/special-issues/special-issue-2024-vol-13/ https://doi.org/10.24191/jmeche.v13i1.3754 |
_version_ |
1817848759511416832 |
score |
13.223943 |