The Effect of Position and Different Size of Radial Hole on Performance of Cannulated Pedicle Screw
Cannulated Pedicle Screw (CPS) is the instrumentation that has been introduced in Biomedical field in order to stabilize the spine structure and also to enhance the pullout strength in osteoporotic patient. The cement is injected through the CPS which is flow from the screw head to the radial hole...
Saved in:
Main Authors: | , , , , , |
---|---|
Format: | Conference or Workshop Item |
Language: | English |
Published: |
EDP Sciences
2017
|
Subjects: | |
Online Access: | http://umpir.ump.edu.my/id/eprint/14261/13/fkm-2017-masayu-The%20Effect%20of%20Position%20and%20Different%20Size.pdf http://umpir.ump.edu.my/id/eprint/14261/ https://doi.org/10.1051/matecconf/201710813001 |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
Summary: | Cannulated Pedicle Screw (CPS) is the instrumentation that has been introduced in Biomedical field in order to stabilize the spine structure and also to enhance the pullout strength in osteoporotic patient. The cement is
injected through the CPS which is flow from the screw head to the radial hole and then distributed around the CPS to
give higher performance of pullout strength. However, the fracture still reported to happened around the radial hole.
This paper is aim to focus on the effect of the position and different diameter of radial hole on the CPS. Three designs were constructed using SolidWork software where two of it are the current models whereas the third was the
proposed design which the radial holes were position differently (skewed to each other) and its size is smaller than the current model, 1.0 mm and 1.5mm respectively. The software used for Finite Element Analysis (FEA) is Ansys
workbench 16.0. The Finite Element Models were verified with FEA result of the previous research. The FEA results
of torsional stress for current and proposed design were then compared to obtain the maximum torque that both designs are able to resist before fracture. It can be concluded that the performance of CPS can be improved by the
proposed design which is 2 radial hole skewed to each other and smaller radial hole diameter (1.0mm) than the current cannulated pedicle screw (1.5mm).
|
---|