Failure analysis of retrieved UHMWPE tibial insert in total knee replacement
This study involves the failure analysis of an ultra high molecular polyethylene (UHMWPE) tibial insert from Apollo Total Knee System, which was removed after 10 years of service from 70 years old female patient. The tibial insert was investigated by using a stereoscope, scanning electron microscope...