Validity of selected White Blood Cells (WBC) differential flags of sysmex XE-2100 in Hospital Selayang / Nur Nazihah Md. Yusof

Automated white blood cells (WBC) differential counts are widely accepted in routine practice, as the accuracy and clinical usefulness have been validated in numerous studies. However, many laboratories still reflexively perform manual WBC differential counts based solely on instrument “flags”. Haem...

全面介绍

Saved in:
书目详细资料
主要作者: Md. Yusof, Nur Nazihah
格式: Thesis
语言:English
出版: 2020
主题:
在线阅读:https://ir.uitm.edu.my/id/eprint/89036/1/89036.pdf
https://ir.uitm.edu.my/id/eprint/89036/
标签: 添加标签
没有标签, 成为第一个标记此记录!
id my.uitm.ir.89036
record_format eprints
spelling my.uitm.ir.890362024-01-09T04:06:52Z https://ir.uitm.edu.my/id/eprint/89036/ Validity of selected White Blood Cells (WBC) differential flags of sysmex XE-2100 in Hospital Selayang / Nur Nazihah Md. Yusof Md. Yusof, Nur Nazihah Immunology Automated white blood cells (WBC) differential counts are widely accepted in routine practice, as the accuracy and clinical usefulness have been validated in numerous studies. However, many laboratories still reflexively perform manual WBC differential counts based solely on instrument “flags”. Haematology laboratory of Selayang Hospital has been using a set of criteria for peripheral blood film (PBF) review, which are adapted from international consensus guidelines (2005) but with modification of several parameters, established by our expert opinion by taking into account our local requirements, and our criteria do not include the analyser’s flags. In this study, we determined the validity of selected 5 most common WBC differential related flags in automated full blood count (FBC) analyser Sysmex XE-2100 by correlating the flags with PBF to verify the safety of our PBF review criteria. Overall, 422 flagged samples were selected during the period of study. The results of both automated and manual counting and scanning of the samples were carefully studied and compared. A total of 538 flags were identified, which consist of 216 atypical lymphocyte flags, 196 blasts flags, 48 WBC AbN scattergram flags, 39 NRBC flags, and 39 monocytosis flags. Qualitatively, there was statistically significant towards negative PBF findings for blasts flags in both age group, and monocytosis flags in the children sample, while the flag of atypical lymphocyte in the children’s age group was statistically significant towards positive findings. Quantitatively, there was a statistically significant difference in the mean of automated and manual monocyte counts in both age groups. The underlying disorders of all these true positive and false-positive flags are heterogeneously varied and they were rather inconsistent and inefficient with low specificity. Based on these findings, we agreed to ignore all the flags and report the FBC using the automated count. Although the atypical lymphocyte flag of the paediatric group was statistically significant towards positive PBF findings, we also ignored this flag as the results did not greatly affect the clinical management of the patients. Facing over-flagging of specimens from an automated FBC analyser is a problem in the haematology laboratory, as a flag requires a time-consuming microscopic review of the specimen, which increases the workload of the laboratory personnel, while they constantly under pressure to provide accurate test results to clinicians promptly. Our goal was to develop a system that directed the use of automated and manual WBC differential assessment towards the greatest clinical gain and increased the efficiency of laboratory operations and enabling laboratory personnel to concentrate on tasks that are no achievable through automated methods. Therefore, in the background of our local patient population, we proposed that there is no clinical rationale for reflex manual PBF examination based on instrument flags, and our current criteria for blood smear review are adequate and safe. 2020 Thesis NonPeerReviewed text en https://ir.uitm.edu.my/id/eprint/89036/1/89036.pdf Validity of selected White Blood Cells (WBC) differential flags of sysmex XE-2100 in Hospital Selayang / Nur Nazihah Md. Yusof. (2020) Masters thesis, thesis, Universiti Teknologi MARA (Kampus Sg. Buloh).
institution Universiti Teknologi Mara
building Tun Abdul Razak Library
collection Institutional Repository
continent Asia
country Malaysia
content_provider Universiti Teknologi Mara
content_source UiTM Institutional Repository
url_provider http://ir.uitm.edu.my/
language English
topic Immunology
spellingShingle Immunology
Md. Yusof, Nur Nazihah
Validity of selected White Blood Cells (WBC) differential flags of sysmex XE-2100 in Hospital Selayang / Nur Nazihah Md. Yusof
description Automated white blood cells (WBC) differential counts are widely accepted in routine practice, as the accuracy and clinical usefulness have been validated in numerous studies. However, many laboratories still reflexively perform manual WBC differential counts based solely on instrument “flags”. Haematology laboratory of Selayang Hospital has been using a set of criteria for peripheral blood film (PBF) review, which are adapted from international consensus guidelines (2005) but with modification of several parameters, established by our expert opinion by taking into account our local requirements, and our criteria do not include the analyser’s flags. In this study, we determined the validity of selected 5 most common WBC differential related flags in automated full blood count (FBC) analyser Sysmex XE-2100 by correlating the flags with PBF to verify the safety of our PBF review criteria. Overall, 422 flagged samples were selected during the period of study. The results of both automated and manual counting and scanning of the samples were carefully studied and compared. A total of 538 flags were identified, which consist of 216 atypical lymphocyte flags, 196 blasts flags, 48 WBC AbN scattergram flags, 39 NRBC flags, and 39 monocytosis flags. Qualitatively, there was statistically significant towards negative PBF findings for blasts flags in both age group, and monocytosis flags in the children sample, while the flag of atypical lymphocyte in the children’s age group was statistically significant towards positive findings. Quantitatively, there was a statistically significant difference in the mean of automated and manual monocyte counts in both age groups. The underlying disorders of all these true positive and false-positive flags are heterogeneously varied and they were rather inconsistent and inefficient with low specificity. Based on these findings, we agreed to ignore all the flags and report the FBC using the automated count. Although the atypical lymphocyte flag of the paediatric group was statistically significant towards positive PBF findings, we also ignored this flag as the results did not greatly affect the clinical management of the patients. Facing over-flagging of specimens from an automated FBC analyser is a problem in the haematology laboratory, as a flag requires a time-consuming microscopic review of the specimen, which increases the workload of the laboratory personnel, while they constantly under pressure to provide accurate test results to clinicians promptly. Our goal was to develop a system that directed the use of automated and manual WBC differential assessment towards the greatest clinical gain and increased the efficiency of laboratory operations and enabling laboratory personnel to concentrate on tasks that are no achievable through automated methods. Therefore, in the background of our local patient population, we proposed that there is no clinical rationale for reflex manual PBF examination based on instrument flags, and our current criteria for blood smear review are adequate and safe.
format Thesis
author Md. Yusof, Nur Nazihah
author_facet Md. Yusof, Nur Nazihah
author_sort Md. Yusof, Nur Nazihah
title Validity of selected White Blood Cells (WBC) differential flags of sysmex XE-2100 in Hospital Selayang / Nur Nazihah Md. Yusof
title_short Validity of selected White Blood Cells (WBC) differential flags of sysmex XE-2100 in Hospital Selayang / Nur Nazihah Md. Yusof
title_full Validity of selected White Blood Cells (WBC) differential flags of sysmex XE-2100 in Hospital Selayang / Nur Nazihah Md. Yusof
title_fullStr Validity of selected White Blood Cells (WBC) differential flags of sysmex XE-2100 in Hospital Selayang / Nur Nazihah Md. Yusof
title_full_unstemmed Validity of selected White Blood Cells (WBC) differential flags of sysmex XE-2100 in Hospital Selayang / Nur Nazihah Md. Yusof
title_sort validity of selected white blood cells (wbc) differential flags of sysmex xe-2100 in hospital selayang / nur nazihah md. yusof
publishDate 2020
url https://ir.uitm.edu.my/id/eprint/89036/1/89036.pdf
https://ir.uitm.edu.my/id/eprint/89036/
_version_ 1787585937316773888
score 13.251813