Analysis of an Improved Hybrid Stem Design for Total Hip Replacement (THR) / Solehuddin Shuib ... [et al.]

This paper proposed an improvement in design and analysis of hybrid prosthesis in total hip replacement. A hip implant is a method for an artificial hip replacement which is made of a material that is different and is selected depending on the study in terms of strength. A hip implant is often focus...

Full description

Saved in:
Bibliographic Details
Main Authors: Shuib, Solehuddin, Ismail, Nur Faiqa, Yahya, Muhd Azman, Ahmed Shokri, Amran
Format: Article
Language:en
Published: Faculty of Mechanical Engineering Universiti Teknologi MARA (UiTM) 2018
Subjects:
Online Access:https://ir.uitm.edu.my/id/eprint/40604/1/40604.pdf
https://ir.uitm.edu.my/id/eprint/40604/
Tags: Add Tag
No Tags, Be the first to tag this record!
_version_ 1833066454361047040
author Shuib, Solehuddin
Ismail, Nur Faiqa
Yahya, Muhd Azman
Ahmed Shokri, Amran
author_facet Shuib, Solehuddin
Ismail, Nur Faiqa
Yahya, Muhd Azman
Ahmed Shokri, Amran
author_sort Shuib, Solehuddin
building Tun Abdul Razak Library
collection Institutional Repository
content_provider Universiti Teknologi Mara
content_source UiTM Institutional Repository
continent Asia
country Malaysia
description This paper proposed an improvement in design and analysis of hybrid prosthesis in total hip replacement. A hip implant is a method for an artificial hip replacement which is made of a material that is different and is selected depending on the study in terms of strength. A hip implant is often focused on the damage to the hip joint. There are various factors which involve damage to the hip such as osteoarthritis, stress shielding, etc. Referring to the various types of total hip replacement products, several factors have been taken to differentiate between each type including the material, design and performance of the product. This project focuses on the stem of total hip replacement. By using computer aid design (CAD) software Solidworks 2013, the stem was redesigned with improvement and underwent the analysis using ANSYS 15.0 to analyses the behaviour which include the total deformation and equivalent (Von-Mises) stress.Higher stress shielding were found at the neck of the hip implant.The three major design types of the hip stems were compared in this study, and the design three (3) hip implant is the most stable among the others after being simulated using different activities.
format Article
id my.uitm.ir-40604
institution Universiti Teknologi Mara
language en
publishDate 2018
publisher Faculty of Mechanical Engineering Universiti Teknologi MARA (UiTM)
record_format eprints
spelling my.uitm.ir-406042021-01-18T01:40:54Z https://ir.uitm.edu.my/id/eprint/40604/ Analysis of an Improved Hybrid Stem Design for Total Hip Replacement (THR) / Solehuddin Shuib ... [et al.] jmeche Shuib, Solehuddin Ismail, Nur Faiqa Yahya, Muhd Azman Ahmed Shokri, Amran TJ Mechanical engineering and machinery This paper proposed an improvement in design and analysis of hybrid prosthesis in total hip replacement. A hip implant is a method for an artificial hip replacement which is made of a material that is different and is selected depending on the study in terms of strength. A hip implant is often focused on the damage to the hip joint. There are various factors which involve damage to the hip such as osteoarthritis, stress shielding, etc. Referring to the various types of total hip replacement products, several factors have been taken to differentiate between each type including the material, design and performance of the product. This project focuses on the stem of total hip replacement. By using computer aid design (CAD) software Solidworks 2013, the stem was redesigned with improvement and underwent the analysis using ANSYS 15.0 to analyses the behaviour which include the total deformation and equivalent (Von-Mises) stress.Higher stress shielding were found at the neck of the hip implant.The three major design types of the hip stems were compared in this study, and the design three (3) hip implant is the most stable among the others after being simulated using different activities. Faculty of Mechanical Engineering Universiti Teknologi MARA (UiTM) 2018 Article PeerReviewed text en https://ir.uitm.edu.my/id/eprint/40604/1/40604.pdf Analysis of an Improved Hybrid Stem Design for Total Hip Replacement (THR) / Solehuddin Shuib ... [et al.]. (2018) Journal of Mechanical Engineering (JMechE) <https://ir.uitm.edu.my/view/publication/Journal_of_Mechanical_Engineering_=28JMechE=29/>, SI 5 (5). pp. 205-2015. ISSN 18235514
spellingShingle TJ Mechanical engineering and machinery
Shuib, Solehuddin
Ismail, Nur Faiqa
Yahya, Muhd Azman
Ahmed Shokri, Amran
Analysis of an Improved Hybrid Stem Design for Total Hip Replacement (THR) / Solehuddin Shuib ... [et al.]
title Analysis of an Improved Hybrid Stem Design for Total Hip Replacement (THR) / Solehuddin Shuib ... [et al.]
title_full Analysis of an Improved Hybrid Stem Design for Total Hip Replacement (THR) / Solehuddin Shuib ... [et al.]
title_fullStr Analysis of an Improved Hybrid Stem Design for Total Hip Replacement (THR) / Solehuddin Shuib ... [et al.]
title_full_unstemmed Analysis of an Improved Hybrid Stem Design for Total Hip Replacement (THR) / Solehuddin Shuib ... [et al.]
title_short Analysis of an Improved Hybrid Stem Design for Total Hip Replacement (THR) / Solehuddin Shuib ... [et al.]
title_sort analysis of an improved hybrid stem design for total hip replacement (thr) / solehuddin shuib ... [et al.]
topic TJ Mechanical engineering and machinery
url https://ir.uitm.edu.my/id/eprint/40604/1/40604.pdf
https://ir.uitm.edu.my/id/eprint/40604/
url_provider http://ir.uitm.edu.my/